Watch one of our OGT experts demonstrate the FISH protocol for our CytoCell® haematology FISH probes. Always refer to the instructions for use supplied with the product. Further validation is required for use in a diagnostic setting.
Watch quick video tips and tricks from our OGT experts for all stages of the CytoCell haematology FISH protocol.
Try to use just one type of immersion oil in your laboratory; different types of oil may be immiscible. Immersion oil of one type remaining on a microscope objective may result in a ‘milky’ appearance and a reduction of light transmission when it comes in contact with a slide with a different type of immersion oil.
All our probe products are optimised to make them suitable for FISH. We generally cut all our probes into fragments between 200 and 2000bp depending on the product. The size of the signal depends on the length of the DNA sequence recognised by the probe when all fragments recognise their target.
Shipping temperatures are dependent on both the product and the destination.
We would suggest that you dispose of your unwanted probes by throwing them in the general lab waste; treating the probes as you would any other low-grade hazardous waste. Please note that the probes should not be flushed down a drain.
The number provided on the kit is the kit lot number. The individual component(s) in the kit will also have their own lot numbers specific for the manufacturing of those products.
Figure 1. Kit Lot Number
Figure 2. Reagent Lot Number
Please inform CytoCell technical support if you suspect that your probe is showing signs of light damage. Our manufacturing team maintains a probe archive which can be tested internally. The common characteristics of light damaged probes are nonspecific probe binding and high autofluorescence.
To mitigate the effects of light damage, we recommend the following:
Example 1 - Light damaged probe
Example 2 - Light damaged probe
The expiry date on the Certificates of Analysis (CoA) and DAPI is displayed in YYYY-MM-DD format on both the component tube and the outer kit label.
Figure 1. DAPI tube label
The expiry date of all other catalogue products is displayed in YYYY-MM format. Please note this extends to the last day of the month, regardless of date of manufacture on the CoA.
Figure 2. Probe tube label
There should not be a mismatch between the expiry date on the probe label and the label on the outer kit. If you have checked the information above and are still concerned that you may have a mismatch in expiry dates, please contact our support team.
CytoCell probes that begin with the catalogue classification LPH correspond to haematology FISH probes and those that begin with LPS correspond to solid tumour/haematopathology FISH probes.
The off-label use of probes is not recommended by CytoCell, the recommended protocol should be followed first to establish expected probe brightness and hybridisation efficiency before deviations from the protocol are considered. CytoCell can only guarantee results on the intended sample type.
Haematology probes have been optimised for use on 3:1 methanol/acetic acid fixed peripheral blood or 3:1 methanol/acetic acid fixed bone marrow.
Solid tumour probes have been validated for use on formalin-fixed paraffin-embedded (FFPE) tissue.
Haemato-pathology probes are validated for dual use on 3:1 methanol/acetic acid fixed bone marrow, 3:1 methanol/acetic acid fixed peripheral blood and FFPE tissue.
Pepsin treatment always improves FISH results, regardless of cell type used -peripheral blood, bone marrow, amniocytes, sputum, etc. CytoCell does not usually recommend pepsin treatment for every product because it is time consuming and requires additional costly reagents and for 'regular' cytogenetics preparations the benefits do not warrant the extra work.
We especially suggest that pepsin treatment be used for late gestational age samples because we have noticed a marked improvement certainly worth the additional effort. This does not mean that pre-treatment is essential or only required for late gestational samples; however, it can be beneficial as some people find it difficult to analyse FISH results without pre-treatment.
If you already routinely perform pepsin digestion for all samples then maintain this practice as it will produce the best results.
A solution of 3:1 Methanol/Acetic Acid (Carnoy’s) should be made freshly prepared, stored at -20°C and discarded after a day.
Try pre-fixing your bone marrow or peripheral blood samples by slowly adding ice-cold fix with agitation, immediately after the hypotonic treatment, to improve your cell preparations.
All labs should validate their own FISH protocol when working with peripheral blood smear preparations. As a suggestion, slides can be immersed in fixation solution from 10-30 mins, apply 10μl probe (or more if required, depending on the coverslip used), denature at 80°C for 5 mins* then follow the recommended CytoCell protocol.
*Please note that this has not been validated internally but based on user feedback.
The optimal conditions for spreading are: 25°C and 50% relative humidity (RH).
Recommendations:
In general, baking slides using a hotplate and/or ageing will produce weaker, more variable results. This tends to be part of older FISH protocols.
CytoCell probes (aside from Sub-telomeric and enumeration probes) are supplied ready to use with probe and hybridisation solution pre-mixed. This mix has been developed and optimised for the intended purpose using 10μl of probe and a 24mm x 24mm coverslip. For this reason, we do not recommend dilution or reduction in volume. Both can produce weaker, more variable probe signals.
Please see the example images below for guidance:
Example 1: IGH Plus Breakapart (LPH 070) using 10ul of probe.
Example 2: IGH Plus Breakapart (LPH 070) using 1ul of probe.
Example 3: IGH Plus Breakapart (LPH 070) using 5ul probe + 5ul Cytocell Hybridisation A (HA500L)
Recommendations:
Example 1: P53 (LPH 017) sealed with a glass coverslip and Weldite glue
Example 2: P53 (LPH 017) sealed with Parafilm
Example 3: P53 (LPH 017) sealed with Cling Film
Example 4: P53 (LPH 017) not sealed with glue.
Ensure pipette tips are long enough so that when the tip reaches the bottom of the vial, the pipette itself doesn’t come into contact. Pipettes entering a probe vial may carry over probe reagent from one vial to the next resulting in cross contamination.
It is vital. The probes which are specific for repeat elements of the human genome are less susceptible to over-denaturation than those specific for single sequences of DNA. They can tolerate much more loss in activity as there are more repeat units to which the probes can hybridise. However, one must also consider the sample. This is made of DNA too, and will lose significant levels of morphology if overdenatured. It is always good practice to take good care when denaturing any FISH product to ensure good signals and good sample morphology.
Figure 1: Example of overdenaturation using Subtelomere 17p and 17q probes with a denaturation of 95°C for 2 mins on cultured lymphocytes.
Figure 2: Example of undenaturation using the MDM2 probe, with a denaturation of 60°C for 2 mins on a skin biopsy sample.
This is a common problem particularly with older or well-used hybridisers, where temperature discrepancies are noticeable on the bottom and top of the heating plate. It is not uncommon to have a position that reads at 60°C denaturation despite the program settings stating 75°C. We recommend regular calibration of hybridisers.
This is likely to be caused by low humidity during the hybridisation step. The FITC (green) fluorophore is more sensitive to these changes and can produce loss of or drop out of control probe signals. The optimal humidity levels for efficient probe hybridisation are 65-70%; however, we would recommend 95-100% humidity to minimise the risk of weak control signals.
Post-hybridisation washing after FISH hybridisation is necessary to aid the removal of non-specific interactions between the probe and undesirable regions of the genome, thus allowing greater probe specificity.
The buffers used in post-hybridisation washing are SSC based and therefore provide positively charged sodium ions in solution. These ions counteract the repulsive negative force between the DNA backbones of both the probe and target. Too much SSC in the buffer will produce a poor washing effect of low stringency and too little SSC (or just water) will tend to wash all probe away from the sample due to high stringency. Temperature and pH also influence the washing effect; increasing the temperature increases the stringency and the pH determines the availability of the positive ions.
The inclusion of TWEEN 20 detergent decreases background staining and enhances the spreading of the reagents in the wash buffers.
Post hybridisation washes at CytoCell are optimised for the FISH procedure and have been formulated using the principles outlined above.
An additional tip would be to periodically wash solution jars and ensure adequate detergent removal can help reduce background issues. Also, the use of filtered pipette tips can reduce intermittent background issues from debris being expelled onto FISH slides.
Figure 1: Example of high stringency >73°C, <0.4xSSC, pH differs from 7.
Figure 2: Example of low stringency <71°C, >0.4xSSC, pH differs from 7.
Once the DAPI counterstain and coverslip have been applied to the slide, it is recommended that users seal the coverslip with varnish. This prevents excessive drying out and maintains the probe signal strength. For short term storage (0-2 weeks), the slides should be covered and stored at room temperature. For long term storage, the slides should be stored 2-8°C and covered.
Currently, there are two different triple filters commonly used in laboratories, one which is specific for TRITC/FITC/DAPI and one which allows visualisation of Texas Red/FITC/DAPI. The first one is easily distinguished by the green light that comes from the objective on the microscope, the other gives a white/pale yellow colour. The only filter that allows simultaneous visualisation of CytoCell's green, red and blue dyes is a Texas Red/FITC/DAPI filter.
Both types of filter can be used, but using the Spectrum Red will give brighter signals for Texas Red probes and using the Spectrum Orange will give brighter signals for our orange fluorophore.
Depending on the manufacturer’s recommendations, fluorescent filters typically need to be replaced every 2-4 years. Over time and use, the filter will begin to degrade. The damaged emission/excitation filters will show a mottled appearance, which begins on the edges and will eventually completely obscure the filter. The effects may not be noticeable on a day to day basis, but will manifest as higher than normal background and weaker signals.
Please contact CytoCell technical support for further advice and guidance if you suspect filter problems.
Depending on the type of bulb used, the shelf life can vary from 200-3000 hours. Please verify with your supplier. The older the bulb, the dimmer the fluorescence signals become, and the greater the chance of the bulb exploding. This is dangerous as mercury vapour is highly toxic.