Watch our video for the denaturation and hybridization steps of the CytoCell® hematology FISH protocol. Study the Package Insert (Instructions for Use, IFU) carefully before using the protocol depicted in this video. This video does not replace the content from the Package Insert. The Package Insert can be found in the product packaging of each FISH Probe Kit and in the Product Documentation section of each FISH Probe Kit on the website. Refer to the Package Insert for warnings, precautions, storage and handling.
It is vital. The probes which are specific for repeat elements of the human genome are less susceptible to over-denaturation than those specific for single sequences of DNA. They can tolerate much more loss in activity as there are more repeat units to which the probes can hybridize. However, one must also consider the sample. This is made of DNA too, and will lose significant levels of morphology if overdenatured. It is always good practice to take good care when denaturing any FISH product to ensure good signals and good sample morphology.
Figure 1: Example of overdenaturation using Subtelomere 17p and 17q probes with a denaturation of 95°C for 2 mins on cultured lymphocytes.
Figure 2: Example of undenaturation using the MDM2 probe, with a denaturation of 60°C for 2 mins on a skin biopsy sample.
This is a common problem particularly with older or well-used hybridizers, where temperature discrepancies are noticeable on the bottom and top of the heating plate. It is not uncommon to have a position that reads at 60°C denaturation despite the program settings stating 75°C. We recommend regular calibration of hybridizers.
This is likely to be caused by low humidity during the hybridization step. The FITC (green) fluorophore is more sensitive to these changes and can produce loss of or drop out of control probe signals. The optimal humidity levels for efficient probe hybridization are 65-70%; however, we would recommend 95-100% humidity to minimize the risk of weak control signals.